Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00190-1
Product Name/Description Tri-Flex II Bed
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 26/02/2014
Responsible Entity Prius Healthcare Solutions Pty Ltd
Reason/Issue CPR quick release mechanism not functional. Wire pulls out of guide slot. Handle to activate is a push instead of a pull or squeeze grip. Handle of mechanism comes in contact with bed frame when head of bed is lowered causing bed frame damage.
Recall Action Recall for Product Correction
Recall Action Instructions Prius Healthcare is completing an update on all affected units.
Contact Information 02 8845 3803 - Prius Healthcare Solutions