Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00174-1
Product Name/Description OPMI Pentero surgical microscope system (PC 2.1)

Manufactured from December 8, 2006 to October 5, 2010

Serial Numbers: 6631402471 to 6631461774

ARTG Number: 187127
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 20/02/2014
Responsible Entity Carl Zeiss Pty Ltd
Reason/Issue Based on post-market surveillance of hardware system performance, it has been identified that system components may cause, in rare situations, a blocking of the magnetic brakes or a system freeze/ system restart during operation potentially prolonging surgery.
Recall Action Recall for Product Correction
Recall Action Instructions Carl Zeiss is providing a software and hardware update to ensure the malfunction does not occur. An enhanced long-term software and hardware maintenance of the OPMI Pentero system procedure is also included in the maintenance package. This action has been closed-out on 28/01/2016