Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00167-1
Product Name/Description FreeStyle Papillon Mini Blood Glucose Monitoring System

List Number: 70728.01

ARTG Number: 126412 (cancelled by the sponsor)
Recall Action Level Consumer
Recall Action Classification Class I
Recall Action Commencement Date 20/02/2014
Responsible Entity Abbott Australasia Pty Ltd T/A Abbott Diabetes Care
Reason/Issue Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or give blood glucose results which could affect clinical outcome or have significant medical risk when used in conjunction with any FreeStyle test strip lot within expiry.br /br /This is an update to the recall initiated in November 2013 on certain lots of FreeStyle Blood Glucose Test Strips (RC-2013-RN-01228-1). The FreeStyle Papillion Mini blood glucose monitoring system has not been supplied in Australia since 2009.br /br /Further inforamtion on this recall can be found at http://www.tga.gov.au/safety/alerts-device-freestyle-strips-papillon-mini-140304.htm
Recall Action Recall
Recall Action Instructions Abbott is providing consumers who are still the possession of of a FreeStyle Papillion Mini blood glucose meter with a replacement FreeStyle branded meter.
Contact Information 1800 801 478 - Abbott Diabetes Care Customer Service