Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00166-1
Product Name/Description Edi Catheter (nasogastric feeding tube)

Edi Catheter,6Fr,49cm,PHT free 5pcs/pkg
Article Number: 6685775

Edi Catheter, 6Fr, 50cm, PHT free 5pcs/pkg
Article Number: 6685777

Edi Catheter,8Fr,100cm,PHT free 5pcs/pkg
Article Number: 6685780

Edi Catheter,8Fr,125cm,PHT free 5pcs/pkg
Article Number: 6685782

Edi Catheter,12Fr,125cm,PHT free 5pcs/pkg
Article Number: 6685787

Edi Catheter,16Fr,125cm,PHT free 5pcs/pkg
Article Number: 6685790
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 26/02/2014
Responsible Entity Maquet Australia Pty Ltd
Reason/Issue Following the report of an Edi Catheter disintegrating in a patient after 17 days of use Maquet has decided to provide the Edi Catheter with its own user's manual. The Edi Catheter IFU states that the catheter can only be in use for a maximum of 5 days.
Recall Action Recall for Product Correction
Recall Action Instructions Maquet is advising end users that the instructions for use will be provided with the Edi catheter and reiterating that the device can only be in use for up to 5 days.