TGA Recall Reference RC-2014-RN-00165-1
Product Name/Description PS500 External Power Supply Unit used with Dräger V Series Ventilators

Models: Infinity V500; Babylog VN500; Evita V300

ARTG Number: 158739
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 18/02/2014
Responsible Entity Draeger Medical Australia Pty Ltd
Reason/Issue The batteries installed in the PS500 deplete much earlier than expected although the battery indicator showed a sufficiently charged battery. In some cases neither the “Battery Low” nor the “Battery depleted” alarm was given, when the remaining battery capacity fell below 10%. Available data indicates that the charged battery will last for at least 3 minutes. In case of a main power loss, this 3 minute battery back-up should typically be sufficient until the emergency main power supply is re-established.
Recall Action Recall for Product Correction
Recall Action Instructions Draeger is advising users to only use the affected device for transport if absolutely necessary and users should not rely on the battery status indicator.
Draeger is recommending users have a breathing bag available for manual ventilation as per the Instructions for Use (IFU).
Draeger is advising that if the power fail alarm occurs during transport, provide manual ventilation and immediately connect the ventilator to wall power to resume ventilation.
Draeger is advising that the batteries on all affected devices will be replaced as an interim correction, until a permanent fix can be developed.