Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00155-1
Product Name/Description Calibrator BRAHMS Free beta-hCG KRYPTOR CAL. An in vitro diagnostic medical device (IVD)

Lot Number: 09045
Expiry Date: 16 October 2014
Kit Reference Number: CDB80991
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/02/2014
Responsible Entity Thermo Fisher Scientific
Reason/Issue Thermo Fisher Scientific would like to inform that the use of Calibrator BRAHMS Free beta-hCG KRYPTOR CAL may result in a decreased Free beta-hCG concentration for BRAHMS GM KRYPTOR QC as well as for patient samples by approximately 4%. This may lead to incorrect risk estimation of chromosomal anomaly of the foetus.
Recall Action Recall
Recall Action Instructions Thermo Fisher Scientific is advising users to stop use of the affected lot number. The affected lot will be replaces by a previous calibrator which may have a shorter expiry date. Thermo Fisher is recommending that results generated with the affected calibrator are re-evaluated in the context of other laboratory and clinical parameters and retest if required.
Contact Information 02 8817 4108 - Thermo Fisher Scientfic