Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00150-1
Product Name/Description The Binding Site SPAPLuS analyser. An in vitro diagnostic medical device (IVD)

All serial numbers with ‘C’ suffix

Catalogue number: TBSIE610

ARTG Number: 201723
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 11/02/2014
Responsible Entity In Vitro Technologies Pty Ltd
Reason/Issue In Vitro Technologies have been notified by the manufacturer that the “Prog Error (overflow) DET Comm” pop-up message has been recently reported on a number of the SPAPLuS analysers with the ‘C’ suffix on the serial number. When the error occurs, analysis will stop and a full shutdown is required to restart testing. In addition to this, it causes results to be lost and the test will be required to be repeated.
Recall Action Recall for Product Correction
Recall Action Instructions In Vitro Technologies are advising users that a build up of static is the cause of the issue. A grounding unit will be installed on all affected units to prevent the error from occurring.
Contact Information 1300 932 136 - In Vitro Technologies