Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00139-1
Product Name/Description Neptune 2 Rover Ultra (Mobile unit used as surgical fluid/smoke waste management system)

Catalogue Number: 0702-002-000
All Serial Numbers

ARTG Number: 164275
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 11/02/2014
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Stryker is implementing upgrades to the Neptune 2 Waste Management System further to the previous recall for product correction in May 2013 (TGA reference: RC-2013-RN-00265-1) where Stryker received reports of serious injury or death in the USA due to the Neptune 2 being connected to a passive chest drainage tube or closed wound drainage device post-operatively.
Recall Action Recall for Product Correction
Recall Action Instructions Stryker is upgrading the Neptune 2 Rover Ultra units with a new top cap, labels and software package to mitigate the potential for incorrect use. Thia action has been closed-out on 28/01/2016.