Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00104-1
Product Name/Description Tytin – Regular Set, Double Spill 50 and 500 capsules (self activating capsules of dental alloy used as a dental restorative)

Tytin – Regular Set, Double Spill – 500 capsules
Part Number: 29949

Tytin – Regular Set, Double Spill – 50 capsules
Part Number:29948

Lot Number: 3-1069

ARTG Number: 99936
Recall Action Level Retail
Recall Action Classification Class III
Recall Action Commencement Date 29/01/2014
Responsible Entity Kerr Australia Pty Ltd
Reason/Issue It was discovered that the product may set 15 seconds faster than specified in the Directions for Use. This set time discrepancy may cause the product to set before there is an opportunity to place it in the tooth cavity.
Recall Action Recall
Recall Action Instructions Kerr is requesting users return affected stock for replacement or credit. This action has been closed-out on 11/02/2016.
Contact Information 1800 643 603 - Kerr Australia Customer Care