Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00098-1
Product Name/Description SPAPLuS Analysers. An in vitro diagnostic medical device (IVD)

Catalogue Number: TBSIE610

Manufactured prior to 22 January 2014
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 24/02/2014
Responsible Entity In Vitro Technologies Pty Ltd
Reason/Issue The black foam on the cell fixing plate of older models of SPAPluS analysers can deteriorate over time which can lead to various faults on the analyser.
Recall Action Recall for Product Correction
Recall Action Instructions In Vitro Technologies is providing their customers with the upgraded Cell Fixing Plate.
Contact Information 1300 932 136 - In Vitro Technologies