Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00097-1
Product Name/Description Holding Forceps for SynReam Reaming Rod 2.5 mm (used during surgery to retain an instrument, reduce a fracture, spread an osteotomy, or hold an implant in place)

Lot Numbers: 5923464 and 5923181

ARTG Number: 153950
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 24/01/2014
Responsible Entity Synthes Australia Pty Ltd
Reason/Issue The angle of the locking lever of Holding Forceps from the affected lots has the incorrect angle and may not lock, resulting in the forceps inability to retain the desired position once applied.
Recall Action Recall
Recall Action Instructions Synthes is advising users to inspect stock and quarantine any affected products to return to Synthes.
Contact Information 02 9855 4620 - Synthes Australia