Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00089-1
Product Name/Description Samaritan PDU 400 (automated external defibrillator)

Serial Numbers between 08P00001003 to 11P00007347
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 3/02/2014
Responsible Entity Aero Healthcare
Reason/Issue The software of Samaritan PDU 400 devices may fail to accurately determine the remaining capacity of the device battery. Rather than emitting an alarm warning the user that the device has insufficient battery to deliver therapy the device may shut down. In such circumstances, an affected device may be unable to operate during a sudden cardiac arrest.

This action was undertaken prior to the agreement of the communication by the TGA.
Recall Action Recall
Recall Action Instructions Areo Healthcare provided replacement PDU 400 devices to all affected customers.
Contact Information 1800 628 881 - Aero Healthcare