Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00029-1
Product Name/Description Oxoid Antimicrobial Susceptibility Discs with Vancomycin (VA30). An in vitro diagnostic medical device (IVD)

Product code: CT0058B

Lot Number: 1174497

ARTG Number: 208273
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 16/01/2014
Responsible Entity Thermo Fisher Scientific Australia & New Zealand
Reason/Issue Investigations by the manufacturer have confirmed that cartridges within this batch may contain individual discs with insufficient amount of antibiotic, resulting in smaller than expected amounts of inhibition. Continued use of the affected lot could result in false indication of resistance to vancomycin.
Recall Action Recall
Recall Action Instructions End users are requested to inspect their stock for affected units. Affected units are to be removed from further use. The requirement for the review of patient results should be determined by the laboratory director.
Contact Information 08 8238 9044 - Thermo Fisher Scientific