Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00019-1
Product Name/Description Desktop Pro R7.01 SP2 and Desktop Pro R4.2, R5.0, R5.1, R6.0, R6.1 (Linear accelerator system)

ARTG Number: 111760
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 13/01/2014
Responsible Entity Elekta Pty Ltd
Reason/Issue During the treatment table, gantry and collimator calibration procedures, it is possible to type the incorrect values or invert a value. If the user does not do the quality assurance (QA) procedures after the calibration, it is possible that the user will not find the error made during the calibration. This could result in delivering the treatment with the patient in the incorrect position resulting in clinical mistreatment.
Recall Action Recall for Product Correction
Recall Action Instructions Users are asked to undertake QA procedures in accordance with the instructions for use (IFU) for Integrity, Precise Table and X-ray Volume Imager (XVI). Users will be informed of the mandatory software upgrade upon its release (expected early 2014) which will mitigate the issue. This action has been closed-out on 02/02/2016.
Contact Information 02 8907 1800 - Elekta Pty Ltd