TGA Recall Reference RC-2014-RN-00016-1
Product Name/Description Precision RXI X-ray Imaging Systems equipped with LCD Video Monitor

LCD Video Monitor Model Number: 2385123-2

LCD Video Monitor Serial Number: 00000001386MC2

ARTG Number: 99423
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 9/01/2014
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue GE Healthcare has recently become aware of a potential safety issue involving the overhead LCD Video Monitor Suspension. There is a potential for the locking nut to loosen out of the LCD Video Monitor Suspension yoke. This will not lead to an immediate fall of the suspended monitor since it will be held in place by the remaining bushing and a screw in the absence of the locking-nut. To date, no injury has been reported due to this.
Recall Action Recall for Product Correction
Recall Action Instructions GE Healthcare is advising users that if inappropriate performance or excessive movement of the monitor is observed to immediately contact GE Healthcare. GE Healthcare is correcting all affected systems.