Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00002-1
Product Name/Description Androstenedione assay run on Immulite, Immulite 1000, Immulite 2000, Immulite 2000XPi analysers. An in vitro diagnostic medical device (IVD)

Immulite/Immulite 1000
Catalogue Number: LKA01
Siemens Material Number: 10381158
Lot Number: 409 and above

Immulite 2000/Immulite 2000XPi
Catalogue Number: L2KA02
Siemens Material Number: 10381188
Lot Number: 314 and above

ARTG Number: 179720
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 2/01/2014
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens Healthcare has confirmed over-recovery of samples with concentrations greater than 5.5ng/mL (19.2nmol/L) with the Immulite/Immulite 1000 and Immulite 2000/ Immulite 2000 XPi Androstenedione assays. Siemens has determined that these lots are linear up to a concentration of 5.5ng/mL (19.2nmol/L), instead of 10ng/mL (34.9nmol/L) as stated in the Instructions for Use (IFU). Quality Control materials at concentrations between 5.5ng/mL (19.2nmol/L) and 10ng/mL (34.9nmol/L) will detect this issue.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is advising their customers to use Quality Control materials with at least two levels having concentrations greater that 5.5ng/mL (19.2 nmols/L). Patient values of greater than 5.5ng/mL (19.2 nmols/L) should be verified using an alternate method.
Contact Information 1800 310 300 - Siemens Technical Support Centre