Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00002-1 |
Product Name/Description |
Androstenedione assay run on Immulite, Immulite 1000, Immulite 2000, Immulite 2000XPi analysers. An in vitro diagnostic medical device (IVD)
Immulite/Immulite 1000 Catalogue Number: LKA01 Siemens Material Number: 10381158 Lot Number: 409 and above
Immulite 2000/Immulite 2000XPi Catalogue Number: L2KA02 Siemens Material Number: 10381188 Lot Number: 314 and above
ARTG Number: 179720 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
2/01/2014 |
Responsible Entity |
|
Reason/Issue |
Siemens Healthcare has confirmed over-recovery of samples with concentrations greater than 5.5ng/mL (19.2nmol/L) with the Immulite/Immulite 1000 and Immulite 2000/ Immulite 2000 XPi Androstenedione assays. Siemens has determined that these lots are linear up to a concentration of 5.5ng/mL (19.2nmol/L), instead of 10ng/mL (34.9nmol/L) as stated in the Instructions for Use (IFU). Quality Control materials at concentrations between 5.5ng/mL (19.2nmol/L) and 10ng/mL (34.9nmol/L) will detect this issue. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Siemens is advising their customers to use Quality Control materials with at least two levels having concentrations greater that 5.5ng/mL (19.2 nmols/L). Patient values of greater than 5.5ng/mL (19.2 nmols/L) should be verified using an alternate method. |
Contact Information |
1800 310 300 - Siemens Technical Support Centre |