Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01356-1
Product Name/Description Trident Tritanium Acetabular Window Trials Case

Catalogue Number: 24024020

Lot Number: 032173

ARTG Number: 140892
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 23/12/2013
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Stryker has become aware that some outer trays of the Trident Tritanium Acetabular Window Trials Case have case handles which were installed upside-down, thus blocking 2 of the 4 slots that allow clips to secure the lid to the case. The hazard identified is the potential for the blue wrap to tear as a result of the protruding unsecured clips making the contents unsterile. Note that the plastic tray insert and instrument contents of this tray are not affected by this issue and may continue to be used independently of the outer tray.
Recall Action Recall
Recall Action Instructions Stryker is requesting their customers to inspect and quarantine any units from the affected lot. A Stryker representative will arrange for the return of any affected product.