Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01355-1 |
Product Name/Description |
STERRAD 100S, STERRAD 50 and STERRAD NX Sterilisation Systems
ARTG Number: 123603 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
23/12/2013 |
Responsible Entity |
|
Reason/Issue |
ASP has received customer complaints by healthcare workers reporting odours/smells potentially emanating from the STERRAD Systems. Odours/smells may also be related to normal breakdown of the oil used in the steriliser vacuum pump on the STERRAD 100S ,STERRAD 50 and/or STERRAD NX . The reported odours/smells do not impact sterility of the devices processed within the STERRAD Systems. ASP is investigating this issue to determine the source of the odour/smell and will continue to perform Planned Maintenance in accordance with each system’s schedule. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
If users observe any odours/smells, Johnson and Johnson Medical is requesting them to leave the room as a precaution and discontinue the use of STERRAD Systems until they are serviced. |
Contact Information |
1800 252 194 - Johnson & Johnson Medical Representative |