Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01345-1
Product Name/Description Liaison Biotrin Control Parvovirus B19 IgM

Part Number: 317011

Lot Numbers: 130010X, 130010X/1, 130012X, 130012X/1 & 130012X/2
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 12/03/2014
Responsible Entity DiaSorin Australia Pty Ltd
Reason/Issue Internal investigation confirmed that the Certificates of Analysis of the Liaison Biotrin Control Parvovirus B19 IgM (P/N 317011), lot numbers 130010X, 130010X/1, 130012X, 130012X/1 & 130012X/2, are reporting an incorrect Negative Control range for the LIAISON XL application, when used in combination with Liaison Biotrin Parvovirus B19 IgM (317010) kit lots numbers 129019X & 129019X/1. The Negative Control range for these lots of Liaison Biotrin Control Parvovirus B19 IgM should be 0.0 - 0.5 Index instead of 0.0 - 0.3 Index as previously reported in the incorrect Certificate of Analysis.
Recall Action Recall for Product Correction
Recall Action Instructions DiaSorin has notified their customers of the change to the range of the Liaison Control for Biotrin Parvovirus B19 IgM.
Contact Information 1300 259 835 - DiaSorin Australia