Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01334-1
Product Name/Description Giraffe Stand Alone Resuscitation Systems with Blender, Giraffe Warmer Integrated Resuscitation systems with Blender, Panda Warmer Integrated Resuscitation systems with Blender, Panda Freestanding Warmer Integrated Resuscitation Systems with Blender, Resuscitation System field upgrade kits with Blender

ARTG Number: 139290
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 18/12/2013
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue GE Healthcare has recently become aware of a potential safety issue associated with the loosening of the blender knob shaft on all Resuscitation units. A loose blender knob shaft may potentially affect the accuracy of the blended output of Air/oxygen gas mixture from the unit which could result in hypoxia or hyperoxia.
Recall Action Recall for Product Correction
Recall Action Instructions GE Healthcare is advising users to perform a blender accuracy check as outlined in the service manual to verify the O2 concentration set at 21% and 100% settings are within the recommended range. If the blender accuracy test results do not meet specification, remove from patient use and contact GE Healthcare. If the blender accuracy results pass, the system is suitable for continued use, however GE is recommending that when delivering oxygen a pulse oximeter is used to monitor the patient.
GE Healthcare will be correcting all affected systems as a permanent fix.
Contact Information 1800 659 465 - GE Healthcare National Call Centre