Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01322-1
Product Name/Description Nova StatStrip and StatSensor Blood Monitoring Systems. An in vitro diagnostic medical device (IVD)
(Point of Care System)

ARTG Number: 197482
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 16/12/2013
Responsible Entity Australasian Medical & Scientific Ltd
Reason/Issue This recall action is to alert users that Lithium batteries utilized beyond their intended useful life may swell and/or crack the external case posing a potential safety hazard. In some cases, the swelling may cause the battery door on the meter to incorrectly seat rendering the meter inoperable.
Recall Action Recall for Product Correction
Recall Action Instructions AMSL is advising users that all batteries without an expiration date label will be replaced. AMSL recommends a battery inspection program is implemented to identify and remove batteries from point of use that do not have expiration dates on the battery label.