Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01315-1
Product Name/Description BD BACTED FX - Top Unit and Bottom Unit (used to detect bacteria and fungi in clinical specimens)

BD BACTED FX - Top Unit
Reference Number: 441385
Serial Numbers: FT2899, FT2914, FT2951 & FT3077

BD BACTEC FX - Bottom Unit
Reference Number: 441386
Serial Numbers: FB1724 & FB1756
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/12/2013
Responsible Entity Becton Dickinson Pty Ltd
Reason/Issue BD has determined that some DB BACTEC FX instruments contain racks that may demonstrate variable indicator light intensities. Specifically, the Anonymous vial indicator lights, which should appear yellow, may instead appear green in colour. This has the potential to lead a user of the instrument to accidentally discard an Anonymous vial or the readings associated with an Anonymous vial, before an anonymous vial has completed protocol. The affected instruments continue to function properly and will not report an incorrect result.
Recall Action Recall for Product Correction
Recall Action Instructions BD is providing a software update to correct the issue, which will be available in February 2014. As an interim measure, BD is providing users with work around instructions to follow until the software upgrade is completed.
Contact Information 02 8875 7165 - Becton Dickinson Service