TGA Recall Reference RC-2013-RN-01310-1
Product Name/Description RAPIDLyte Arterial Blood Sampling Line Draw Syringes

3 mL Line Draw, 70 IU, Aspirating, L/L
Reference number: 04180028
Siemens Part Number: 10320296

3 mL Line Draw, 21 IU, Aspirating, L/L
Reference number: 06194484
Siemens Part Number: 10323925

3 mL Line Draw, 70 IU, Aspirating, L/L
Reference number: 06380806
Siemens Part Number: 10490966

3 mL Line Draw, 70 IU, Aspirating, L/S
Reference number: 00355281
Siemens Part Number: 10313293

Multiple lot numbers affected
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 20/12/2013
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens is advising of an issue with RAPIDLyte Arterial Blood syringes when used with Radiometer brand blood gas analysers. The plunger tip in the RAPIDLyte 3mL arterial draw line does not remain stationary when the probe of the analyser extends into the syringe. This can result in difficulties aspirating the sample or the plunger may get pushed out of the syringe.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is advising their customers to cease using RAPIDLyte syringes on Radiometer Blood Gas systems. The syringes can continue to be used on any Siemens Blood Gas Analysers.
Contact Information 1800 310 300 - Siemens Technical Support Centre