Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01294-1 |
Product Name/Description |
Large and Mega Needle Driver Endowrist Instruments for da Vinci Surgical Systems
Large Needle Driver, da Vinci S and da Vinci Si Product Number: INS-420006
Mega Needle Driver, da Vinci S and da Vinci Si Product Number: INS-420194
Large Needle Driver, da Vinci Standard Product Number: INS-400006
Mega Needle Driver, da Vinci Standard Product Number: INS-400194.
Multiple Lot Numbers affected Manufactured between July 2007 and October 2011 ARTG Number: 146826 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
9/12/2013 |
Responsible Entity |
|
Reason/Issue |
Intuitive Surgical has become aware of the potential for detachment of the jaw insert for Large Needle Drivers and Mega needle Drivers. Only Large Needle Driver and Mega Needle Driver instruments manufactured between July 2007 and October 2011 are affected. If a jaw insert detaches during surgery, this could result in a device fragment falling into the patient. |
Recall Action |
Recall |
Recall Action Instructions |
Device Technologies is advising users to quarantine affected devices from use. Affected devices can be returned to Device Technologies. |
Contact Information |
1800 429 551 - Device Technologies Australia |