Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01289-1
Product Name/Description SANGOFIX IV Administration Set and HEIDELBERG EXTENSION TUBING

SANGOFIX IV Administration Set
Article code: 4050193
Batch number: 3H24278T34
ARTG Number: 137615

HEIDELBERG EXTENSION TUBING
Article code: 4052197
Batch number: 3H21278A23
ARTG Number:139265
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 3/12/2013
Responsible Entity B Braun Australia Pty Ltd
Reason/Issue Following reports of leaking from the in-line Safsite injection site, initial investigations have identified that on disconnection of a luer connector used to access the injection site, a resealing issue may occur intermittently. This may result in fluid leaking from the valve.
Recall Action Recall for Product Correction
Recall Action Instructions As an interim corrective action, B.Braun is recommending that end users fit a stand-alone Safsite valve to the in-line injection site. The valve can be used for injection or infusion and does not need to be removed. Further notifications will be undertaken if any additional actions are required based on B. Braun's investigations.
Contact Information 1800 172 182 - B. Braun Customer Service