Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01289-1 |
Product Name/Description |
SANGOFIX IV Administration Set and HEIDELBERG EXTENSION TUBING
SANGOFIX IV Administration Set Article code: 4050193 Batch number: 3H24278T34 ARTG Number: 137615
HEIDELBERG EXTENSION TUBING Article code: 4052197 Batch number: 3H21278A23 ARTG Number:139265 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
3/12/2013 |
Responsible Entity |
|
Reason/Issue |
Following reports of leaking from the in-line Safsite injection site, initial investigations have identified that on disconnection of a luer connector used to access the injection site, a resealing issue may occur intermittently. This may result in fluid leaking from the valve. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
As an interim corrective action, B.Braun is recommending that end users fit a stand-alone Safsite valve to the in-line injection site. The valve can be used for injection or infusion and does not need to be removed. Further notifications will be undertaken if any additional actions are required based on B. Braun's investigations. |
Contact Information |
1800 172 182 - B. Braun Customer Service |