Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01270-1
Product Name/Description VITROS Mixing Cup Arrays on VITROS 250 Chemistry Systems & VITROS 350 Chemistry Systems. An in vitro diagnostic medical device (IVD).

Product Code: 1631779

ARTG Number for VITROS Chemistry Systems:180199
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 2/12/2013
Responsible Entity Ortho-Clinical Diagnostics
Reason/Issue While performing an on-analyser dilution, some customers have reported an increase in EAE condition codes (Sample Metering Dilution Error). The data indicates that the rate of occurrence for EAE codes can vary between 2 and 25% for cups within the affected arrays. When an EAE condition code occurs, the processing for that sample is automatically cancelled by the system.
Ortho-Clinical Diagnostic's investigation into the increased rate of EAE condition codes determined that some of VITROS Mixing Cup Arrays may be out of specification.
Recall Action Recall for Product Correction
Recall Action Instructions Ortho-Clinical Diagnostics is advising their customers to continue using the Mixing Cup arrays as there is currently no alternative available. It is recommended that laboratories perform manual dilutions for samples with low volume rather than using an onboard dilution. This action has been closed-out on 10/02/2016.