Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01269-1 |
Product Name/Description |
DRX Revolution Mobile X-Ray System
All lot numbers affected
ARTG Number: 194055 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
9/12/2013 |
Responsible Entity |
|
Reason/Issue |
The manufacturer, Carestream Health Inc has received reports of the below incidents: - In the USA, the end cover of the DRX Revolution Mobile X-ray System column was dislodged when it was caught on an interfering object whilst the column was descending. - Four occurrences in the USA of a communication failure between PC (host computer) and Communication I/O board (CIB) resulting in the generator not being reset after a successful exposure and image acquisition. This resulted in an exposure initiated by the user using the hand switch completing without the image being captured. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Carestream Health is installing a replacement end cap and software update to correct the issues. |
Contact Information |
02 8962 4914 - Carestream Health Australia |