TGA Recall Reference RC-2013-RN-01257-1
Product Name/Description Calibration Diskette and Assay Data Disks used with VITROS chemistry instruments. An in vitro diagnostic medical device (IVD)

VITROS Chemistry Instruments:
-VITROS 4600
-VITROS 5600
-VITROS 250
-VITROS 350
-VITROS 5,1 FS
-enGen system

VITROS Chemistry Products:
ALT slides -1655281
CK slides -8479396, -8478034
LDH slides -8384489
hsCRP reagent -6801739
THC reagent -6801994
TIBC kit -8867541
METD reagent -6801996
CRBM slides -8892382
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/11/2013
Responsible Entity Ortho-Clinical Diagnostics
Reason/Issue Ortho-Clinical Diagnostics, Inc. (OCD) has determined that the default dilution factors on the Calibration Diskette (Cal Disk) and Assay Data Disks (ADD), DRV 5740 and below, are incorrect. It has also been determined that the 2nd dilution factors contained in the enGen Laboratory Automation System configuration (.gsb) file (Version 3.2.2 & Below) are incorrect.
Recall Action Recall for Product Correction
Recall Action Instructions Ortho-Clincial Diagnostics is providing their customers with Local Cal Disk or ADD version 5741 which contains the correct dilution factors. Customers are also advised to review their LIS data to determine the impact of the revised default dilution factors. For those customers using the enGen system, an Ortho-Clinical Diagnostics representative will assist in the development of revised dilution factors. This action has been closed-out on 12/02/2016.