TGA Recall Reference RC-2013-RN-01256-1
Product Name/Description Leica M525 F50 Surgical Microscope with XY-Unit (surgical light microscope)

Material Number: 10448505

Serial numbers: 260213001, 70613002, 80911001, 110613001, 180712001, 30913001

Manufactured from 10 May 2010 to 18 November 2013

ARTG Number: 137683
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 29/11/2013
Responsible Entity Leica Microsystems Pty Ltd
Reason/Issue Leica Microsystems has identified that in certain activations of the XY, the system can continue to run laterally to its limit (run-on), but will stop if the surgeon activates any button. Run-on of the XY can only occur in the rare situation of rapid consecutive activation on the 45 degrees, 135 degrees, 225 degrees or 315 degrees. In addition, the run-on will stop if any other activation of the XY occurs. The XY run-on is disruptive as the adjustment to the field of view could take up to 24 seconds longer than usual, however there is a low risk to patients.
Recall Action Recall for Product Correction
Recall Action Instructions Leica is providing users with work around instructions to follow as an interim measure. A software upgrade will be implemented as a permanent fix.
Contact Information 02 8870 3500 - Leica Microsystems