Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01255-1
Product Name/Description Patient Data Module (PDM) used with B850, B650, B450, Solar 8000i, MacLab / ComboLab (Multifunction patient monitoring)

All PDMs produced from April 2007 through June 2010

Serial Numbers: SA30713XXXXXX to SA31026XXXXXX

ARTG Number: 118982
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 27/11/2013
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue Fluid can pass through the seals in the top cover and come into contact with energised contacts in the circuit board. This can result in heat buildup, smoke, and discoloration. This heat buildup could potentially result in thermal injury to a patient or a caregiver.
Recall Action Recall for Product Correction
Recall Action Instructions GE Healthcare is advising users to inspect the top cover for signs of heat buildup or discolouration. Remove from use if any evidence of damage is observed. GE is advising users to refrain from spraying fluid onto device. A new cover will be installed on all affected devices as a permanent fix when available. This action has been closed-out on 11/02/2016.
Contact Information 1800 659 465 - GE Healthcare National Call Centre