Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01253-1
Product Name/Description CoreValve AccuTrak Delivery Catheter System (DCS) (Intended to deliver CoreValve Transcatheter Aortic Valve, which is designed to treat severe aortic valve stenosis without open-heart surgery or surgical removal of the native valve)

Models DCS-C4-18Fr and DCS-C4-18Fr-23

Supplied in Australia under the Clinical Trial Notification (CTN) and Authorised Prescriber Schemes
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 26/11/2013
Responsible Entity Medtronic Australasia Pty Ltd
Reason/Issue Medtronic has received thirty-eight (38) reported nose cone separation events (.08 percent of implant procedures); five (5) of these reported events (.01 percent of implant procedures) led to the need for surgical intervention. A thorough review and investigation of the reported events has identified that a majority of these reported events occurred in situations cautioned against in current labelling. In addition, Medtronic has determined that the following situations can lead to nose cone separations:
-Using the CoreValve AccuTrak DCS to retrieve (remove) a partially deployed valve.
- Removing the CoreValve AccuTrak DCS from the patient, after valve deployment, without fully closing the CoreValve AccuTrak DCS capsule.
-Continuing to pull on the CoreValve AccuTrak DCS if increased resistance is felt at the vessel introducer, the introducer’s hemostatic valve, or other structure during CoreValve AccuTrak DCS retrieval.
Recall Action Recall for Product Correction
Recall Action Instructions To reduce the occurrence of nose cone separations, Medtronic is modifying the instructions for use (IFU) and training materials for CoreValve AccuTrak DCS models DCS-C4-18Fr and DCS-C4-18Fr-23. Medtronic is advising that physicians implement the updated IFU recommendations to reduce occurrences of nose cone separation events during the CoreValve implant procedure.