Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01252-1
Product Name/Description Sysmex Automated Blood Coagulation Analyzer CS-2100i and 2000i. An in vitro diagnostic medical device (IVD)

Catalogue Numbers: SMN 10488583 and 10488585

Siemens Reference Number: PH 14-002

ARTG Number: 179939
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 26/11/2013
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue The manufacturer, has become aware about a software bug with Sysmex CS-2100i/2000i Analyzers. When the operator registers an order for micro mode and tries to modify the order, there is a risk of a mismatch between the results and sample ID numbers. Under certain circumstances, measurement results could be assigned to a wrong sample.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is providing users with work around instructions to follow. A software update will be implemented to permanently correct the issue.
Contact Information 1800 310 300 - Siemens Technical Support Center