Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01248-1
Product Name/Description Patient Side Manipulators - da Vinci S, Si, Si-e Surgical System (robotic surgical unit)

Manufactured from October 2012 to April 2013

ARTG Number: 97348
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 27/11/2013
Responsible Entity Device Technologies Australia Pty Ltd
Reason/Issue Intuitive Surgical has become aware that friction within certain Patient Side Manipulators (ie, instrument arms) can interrupt smooth instrument motion. This interruption is felt by the surgeon as resistance in the movement of the mast. In this situation, the instrument can stall momentarily and then suddenly catch up to the master position if the surgeon pushes through the resistance. This resistance is most noticeable when moving the instrument at very slow speeds and during small movements of the masters. This system behaviours is not autonomous instrument motion. The variable function is immediately apparent to the surgeon, and they would exercise surgical judgment to continue use in only situations where the risk is low. However, they would not necessarily be anticipating an interruption of smooth instrument motion, so may not avoid all vulnerable structures.
Recall Action Recall for Product Correction
Recall Action Instructions Device Technologies is testing affected Patient Side Manipulators (instrument arms) and replacing or repairing instrument arms as required. Device Technologies in providing work around instructions for users to follow as an interim measure.
Contact Information 1800 429 551 - Device Technologies Customer Service