Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01246-1
Product Name/Description Modular Finishing Instrument Tray Assembly (used to store, transport and organise instrumentation for the Restoration Modular Instruments Set)

Catalogue Number: 62789970

Lot Number: 039980

ARTG Number: 140892
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/11/2013
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Stryker has become aware that the trays of the affected lot number are missing two of the four necessary cut outs. These cut outs provide clearance to secure the clips on the tray lid to the base. The hazard identified is the potential for the blue wrap to tear as a result of the protruding unsecured clips, making the contents of the tray unsterile. A potential tear would likely be noticed by the surgical staff before the instrument is used as the blue wrap is checked for any defects prior to surgery.
Recall Action Recall
Recall Action Instructions Stryker is advising users to remove any affected trays from use. Any affected trays will the replaced by Stryker.
Contact Information 1800 803 601 - Stryker Australia Quality Assurance