Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01242-1
Product Name/Description STERRAD 100NX (Gas Plasma Steriliser)

Product Number: 10104-002

Affected Serial Numbers: 1042130065, 1042130028, 1042130024, 1042130067, 1042130002, 1042130063, 1042130064

ARTG Number: 123603
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 22/11/2013
Responsible Entity Johnson & Johnson Medical Pty Ltd
Reason/Issue The manufacturer, ASP has determined that welds could loosen on the frame of STERRAD 100NX sterilizers manufactured between October 2012 and June 2013. These welds are used to secure support rails to the top tray assembly for added rigidity. The failure of the broken weld could potentially cause the chassis to deflect concaving downward. This failure does not impact the STERRAD 100NX sterilizer components performance.
Recall Action Recall for Product Correction
Recall Action Instructions A JJM Field Service Engineer will install rivets and brackets to ensure the support rail brackets are firmly attached to the chassis to prevent possible separation.