Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01238-1
Product Name/Description VITROS Chemistry Products K+ Slides. An in vitro diagnostic medical device (IVD)

Catalogue Number: 8157596

Coatings: 0884, 0885
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 21/11/2013
Responsible Entity Ortho-Clinical Diagnostics
Reason/Issue While performing an internal investigation, Ortho Clinical Diagnostics (OCD) determined that the coatings of the VITROS Chemistry Products K+ Slides contain a surfactant that was unintentionally added during the manufacturing process. The affected slides continue to meet intended use and performance specifications, however the long term product performance is unknown.
Recall Action Recall
Recall Action Instructions OCD is requesting their customers to discontinue use and discard all units from the affected lots. Replacement kits will be provided if required.
Contact Information 1800 032 359 - Ortho Clincal Diagnostics Customer Technical Services