Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01237-1 |
Product Name/Description |
Inflation bulbs used with Safescope Sigmoidoscopes
Product Code: SCU-SS220
Batch Number: 061326
ARTG Number: 165977 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
9/12/2013 |
Responsible Entity |
|
Reason/Issue |
Med Chem is recalling affected bulbs following reports of inflation bulbs not compressing. After physical inspections of the fault by both Logikal Health and its supplier, it has been determined that this fault is due to a misshaped component within the valve located on the tubing which extends from the bellows. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Med Chem Surgical is requesting their customers to test any remaining product from the affected batch by applying slight pressure through the packaging. Any bulbs that fail to compress must be returned immediately. |
Contact Information |
02 9699 6460 - Med-Chem Surgical Pty Ltd |