Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01232-1 |
Product Name/Description |
Medfusion 4000 Syringe Infusion Pump
All lot numbers affected
ARTG Number: 122405 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
21/11/2013 |
Responsible Entity |
|
Reason/Issue |
Smiths Medical has identified an issue with electrical shorting at the AC power receptacle. Shorting can occur when saline solution or other conductive fluids (e.g., potassium chloride) enter the AC power receptacle. When this happens, sparking and/ or flames, accompanied by smoke, may appear at the AC power receptacle. The pump and the AC power cord are constructed of flame retardant materials; therefore, the sparks/ flames extinguish within seconds. There is a potential for delay in treatment while a replacement or alternative method of delivery is obtained. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Smiths Medical is advising users to immediately unplug the AC power cord from the power source if there are signs of sparking, flames or smoke emitting from the unit. Smiths Medical is providing work around instructions for users to follow until Smiths Medical develops a permanent fix. |
Contact Information |
1800 654 949 - Smith's Medical Customer Service Department |