Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01217-1
Product Name/Description STERRAD Boosters and Adaptors (used in STERRAD NX steriliser and contains hydrogen peroxide sterilant)

ARTG Number: 123602
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 15/11/2013
Responsible Entity Johnson & Johnson Medical Pty Ltd
Reason/Issue It has been determined that the instructions for use (IFU) for the STERRAD Booster and STERRAD Adaptor include incorrect directions for when a STERRAD Booster and Adapter I, II and III is required to be used. It has also been determined that the current procedures for fitting a STERRAD Booster do not adequately mitigate the risk of improper seal between the scope and the booster or inadequate cleaning of the Adaptor/scope interface.
Recall Action Recall for Product Correction
Recall Action Instructions J&J Medical is advising customers of the correct procedure to follow and providing the updated instructions for use (IFU) to ensure the devices are properly sterilised. A new checklist has also been developed to provide additional verification of the proper use of the STERRAD booster/adapter before and after the STERRAD cycle. This action has been closed-out on 10/02/2016.
Contact Information 1800 252 194 - Johnson & Johnson Medical Customer Service