Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01133-1
Product Name/Description Clinac, Trilogy, Trilogy Tx, Novalis Tx and Unique accelerators, software versions 7.x, 8.x and 9.0. (Linear accelerator system)

Product Codes: H14, H18, H27 and H29

ARTG Number: 116839
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 6/11/2013
Responsible Entity Varian Medical Systems Australasia Pty Ltd
Reason/Issue The manufacturer has received reports in which a user has remotely rotated the gantry into contact with the couch or with patients, in both the manual mode and the automated mode.
Recall Action Recall for Product Correction
Recall Action Instructions Varian is reminding users to ensure patients are observed during manual and automatic gantry movements. Varian is providing a software update to permanently correct the issue. This action has been closed-out on 01/02/2016.
Contact Information 1800 657 036 - Varian Helpdesk