Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01128-1 |
Product Name/Description |
ABL90 FLEX Analyser. An in vitro diagnostic medical device (IVD) (blood gas analyser)
|
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
6/11/2013 |
Responsible Entity |
|
Reason/Issue |
RADIOMETER has recently become aware of a potential issue of negative bias on Sodium results if the ABL90 is not on a horizontal table. When tilted by 10 degrees or more and idle for 30 -45 minutes the magnitude of the effect on sodium results is as large as 10 mM (7%) in the worst case. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Radiometer is advising users to ensure that the ABL90 is placed on a horizontal platform/table. If the analyser is on a roll stand users must ensure that the support plate is aligned horizontally and locked in position. Radiometer will perform alignment of roll stands on request. A detection mechanism will be provided in an upgraded version of the analyser software to eliminate the possibility of error. |
Contact Information |
1800 247 254 - Radiometer Pacific |