Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01128-1
Product Name/Description ABL90 FLEX Analyser. An in vitro diagnostic medical device (IVD)
(blood gas analyser)

Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 6/11/2013
Responsible Entity Radiometer Pacific Pty Ltd
Reason/Issue RADIOMETER has recently become aware of a potential issue of negative bias on Sodium results if the ABL90 is not on a horizontal table. When tilted by 10 degrees or more and idle for 30 -45 minutes the magnitude of the effect on sodium results is as large as 10 mM (7%) in the worst case.
Recall Action Recall for Product Correction
Recall Action Instructions Radiometer is advising users to ensure that the ABL90 is placed on a horizontal platform/table. If the analyser is on a roll stand users must ensure that the support plate is aligned horizontally and locked in position. Radiometer will perform alignment of roll stands on request. A detection mechanism will be provided in an upgraded version of the analyser software to eliminate the possibility of error.
Contact Information 1800 247 254 - Radiometer Pacific