Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01126-1 |
Product Name/Description |
Weinmann Medumat Transport Ventilator
All serial numbers
ARTG Number: 190246 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
1/11/2013 |
Responsible Entity |
|
Reason/Issue |
The manufacturer, Weinmann Emergency, has identified a potential issue that could result in an undefined operating state of the Medumat Transport Ventilator. In some cases an interaction between internal structural components and sensors for internal device monitoring can cause an undefined operating state resulting in the device displaying a "device malfunction" error. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Maquet is advising users to switch the machine off and on again in the case of the error message "device malfunction". This will suspend running ventilation for approximately 30 seconds. Maquet is providing a software update to eliminate incorrect "Device Malfunction" errors. Users should note that any "Device Malfunction" errors after the installation of the software update are valid. |
Contact Information |
07 3339 3936 - Maquet Australia Service Manager |