Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01121-1 |
Product Name/Description |
AxSYM Valproic Acid Reagent (An in vitro diagnostic medical device IVD)
List Number: 7A71.20
Lot Number: 18439M503
ARTG Number: 13401 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
1/11/2013 |
Responsible Entity |
|
Reason/Issue |
Abbott has confirmed that some kits of AxSYM Valproic Acid Reagent lot 18439M503 have the potential to generate invalid calibration, control or sample runs due to Net Intensity Too Low, resulting in AxSYM instrument Error Code 1072. When a valid calibration and/or control/sample run is obtained, patient results generated with reagent lot 18439M503 are acceptable. |
Recall Action |
Recall |
Recall Action Instructions |
Abbott is providing replacement stock for users. Once replacement stock has been received, destroy any remaining units of the affected stock. Abbott is advising that if a valid calibration and/or control/sample run is obtained, patient results are acceptable. |
Contact Information |
1800 816 696 - Abbott Diagnostics Customer Support |