Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01119-1 |
Product Name/Description |
i-STAT PT/INR Cartridge (prothrombin time test used for monitoring patients receiving oral anticoagulation therapy) (An in vitro diagnostic medical device (IVD))
List Number: 3P89-24
Lot Numbers: C103139 to and including C13270A
ARTG Number: 213851 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
1/11/2013 |
Responsible Entity |
|
Reason/Issue |
Abbott Point of Care has determined that the i-STAT PT/INR cartridge lots have the potential to exhibit incorrectly elevated results. Internal studies have demonstrated that i-STAT PT/INR results are elevated by an average of approximately 10% as compared to the International Reference Preparation (rTF/09) in the therapeutic range of 1.8 to 3.0 INR. |
Recall Action |
Recall |
Recall Action Instructions |
Abbott Diagnostic is advising users to destroy unused stock. Any remaining stock will be replaced by Abbott. A review of previous results may be performed at the discretion of the laboratory manager. |
Contact Information |
1800 816 696 - Abbott Diagnostics Customer Support |