Reason/Issue |
Non-Invasive Blood Pressure (NIBP) - Auto-cycling not starting/stopping, NIBP measurement not performed if network time update occurs at same time, infants limits are incorrect. ECG- Wrong ECG bandwidth in E-PSM(P) or E-(P)RE(S)TN module. Changing primary ECG lead selection will not change primary measurement in PDM or TRAM. Central Monitor- A temporarily hidden alarm message will not reactivate if different alarm is acknowledged, audible alarm missing after broken network connection, ECG not shown on iCentral, standby doesn't end, & incorrect labelling on V leads. Ventilation Alarms- Incorrect display of Airway pressure in bed to bed view, spirometry alarm limits can't be changed. General- Alarm off issues in small parameter window, auto-discharge doesn't erase data, invasive pressure lables incorrect. TRAM -Missing service alarm Bed to bed view doesn't show highest ST value in S/5 network, missing remote alarm & AVOA alarm, monitor disconnect alarms not restored. |