Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01099-1 |
Product Name/Description |
VITROS Chemistry Products Na+ Slides and Calibrator Kit 2 (used on VITROS 250/350/5, 1 FS, 4600 Chemistry Systems, VITROS 5600 Integrated System). An in vitro diagnostic medical device (IVD)
VITROS Chemistry Products Na+ Slides Product Code: 8379034 Lot Numbers: GENs 7 to 13 Expiry Dates: 01/11/2013 to 01/01/2015
VITROS Chemistry Products Calibrator Kit 2 Product Code: 1662659 Lot Numbers: 212, 292 Expiry Dates: 31/08/2014, 28/02/2014 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
25/10/2013 |
Responsible Entity |
|
Reason/Issue |
Ortho Clinical Diagnostics (OCD) has conformed that positively biased serum or plasma sample results may occur by up to +1.5mmol/L. |
Recall Action |
Recall |
Recall Action Instructions |
OCD will replace your remaining inventory for VITROSĀ® Na+ Slides calibrated with Lots 212 and 292. It is acceptable to use Lot 212 or 292 to calibrate VITROSĀ® CHOL, Cl-, ECO2, and TRIG Slides until your replacement product arrives. This action has been closed-out on 01/02/2016. |
Contact Information |
1800 032 359 - Customer Technical Service |