Type of Product |
Medical Device |
TGA Recall Reference |
RC-2013-RN-01091-1 |
Product Name/Description |
Novacell Battery Charger supplied with the BLS-1000 Light Source set (light source for endoscopic procedures)
Model number of battery charger: NV-LC02 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
24/10/2013 |
Responsible Entity |
|
Reason/Issue |
AshMed has been made aware of the potential for over-heating and possible burning or bursting of lithium-ion batteries with the affected Vision Sciences battery charger model. When the battery is unintentionally placed by the user in a reversed polarity position for charging, the anode pin on the charger may come into contact with the negative polarity of the battery causing overheating. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
AshMed Pty Ltd is requesting their customer to inspect their Vision Sciences battery charger to determine if they are affected by the issue. Affected units are to be immediately quarantined. AshMed Pty Ltd will arrange for replacement with unaffected chargers. |
Contact Information |
1300 680 898 - Ashmed Pty Ltd |