TGA Recall Reference RC-2013-RN-01077-1
Product Name/Description GSI Viewer, prior to version 2.00-0M on AW and prior to version 2.20-0B in AW Server (Picture archiving and communication system, PACS)

ARTG Number: 161257
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 17/10/2013
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue The GSI Viewer application running on AW/Volume Share or on the AW Server when making a batch and rotating the batch group clockwise 180 degrees (Inferior to Superior) will display images in the Superior orientation. The orientation annotation on the image is correct. The concern is that the patient right/left representation is from the Superior view, which could result in a misreading of the image and misdiagnosis.
Recall Action Recall for Product Correction
Recall Action Instructions GE Healthcare is advising users that the orientation annotation is correct on the image, and to always correlate any information in any 3D reconstruction with the original data. GE Healthcare advises that a software correction will be installed on all affected systems by March 2014.