TGA Recall Reference RC-2013-RN-01065-1
Product Name/Description Vygon Nutrisafe Syringe Feeding Pumps

Product Codes: VY-SN-50C6-A (Orange) and VY-SN-50C6 (Purple)

ARTG Number: 198722
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 28/10/2013
Responsible Entity Device Technologies Australia Pty Ltd
Reason/Issue A calibration fault has been found when using different syringe sizes when using the Orange and Purple Vygon Nutrisafe Syringe Feeding Pumps. The 35ml syringe is not properly recognised by the pump. Additionally, only Channel 2 is correctly calibrated for the Orange Feeding Pump.
Recall Action Recall for Product Correction
Recall Action Instructions Device Technologies is providing users with work around instructions to follow as an interim measure. Device Technologies will contact all users to arrange for the return and correction of affected pumps taking into account operational needs. This action has been closed-out on 11/04/2016.