Type of Product Medical Device
TGA Recall Reference RC-2013-RN-01054-1
Product Name/Description Volumed VP7000 Premium, Volumed VP7000 Classic Plus, Syramed SP6000 Premium and Syramed SP6000 Classic Plus

ARTG Number: 174698, 174697
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/10/2013
Responsible Entity Medevco Pty Ltd
Reason/Issue Medevco identified that in some rare cases the key input line can trigger random key presses. Even though the pumps correctly display the actual settings such as rate and volume infused at all times, random key presses can lead to alarms or undesired manipulations of the set values.

This recall action was carried out prior to approval of the recall strategy by the TGA.
Recall Action Recall for Product Correction
Recall Action Instructions Allied Medical is providing users with a software update to permanently correct the issue.
Contact Information 08 9266 0100 - Allied Medical